FDA Inspection 718287 — USI Orthopaedics, LLC

USI Orthopaedics, LLC — FEI 3007511979

The FDA completed an inspection of USI Orthopaedics, LLC on Apr 6, 2011 in Winter Springs, FL. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 10 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Voluntary Action Indicated (VAI)
End Date
Apr 6, 2011
Fiscal Year
2011
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Winter Springs, FL (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
1471821 CFR 820.30(g)Design validation - Risk analysis
243021 CFR 820.30(b)Design plans - Lack of or inadequate
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
323321 CFR 820.72(a)Calibration, Inspection, etc. Procedures Lack of or Inadequ
341521 CFR 820.22Quality Audit/Reaudit - conducted
407021 CFR 820.30(g)Design validation - documentation
48621 CFR 820.50(a)Evaluation of suppliers, contractors, etc., requirements
54621 CFR 820.75(a)Lack of or inadequate process validation
63021 CFR 803.17Lack of Written MDR Procedures