FDA Inspection 775196 — Genicon, Inc.

Genicon, Inc. — FEI 3002590791

The FDA completed an inspection of Genicon, Inc. on Mar 29, 2012 in Winter Park, FL. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 7 citations were issued during this inspection. Product area: Devices.

Inspection details

Classification
Voluntary Action Indicated (VAI)
End Date
Mar 29, 2012
Fiscal Year
2012
Product Type
Devices
Project Area
Compliance: Devices
Location
Winter Park, FL (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
232721 CFR 820.22Quality audits - Lack of or inadequate procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
319221 CFR 820.30(g)Design validation - user needs and intended uses
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
54621 CFR 820.75(a)Lack of or inadequate process validation

Need the full FDA record for Genicon, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.