Genicon, Inc. FDA Inspection Record
Genicon, Inc. is an FDA-inspected facility (FEI 3002590791) located in Winter Park, FL, US. RegDataLab tracks 5 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Jan 17, 2020: Voluntary Action Indicated (VAI)
- Inspections on record
- 5 (5 VAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 5, latest Sep 17, 2020
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jan 17, 2020 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures for distribution (21 CFR 820.160(a)) Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design verification - Lack of or inadequate procedures (21 CFR 820.30(f)) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) and 11 more |
| Oct 20, 2017 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Documentation (21 CFR 820.100(b)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) Lack of or inadequate process validation (21 CFR 820.75(a)) and 1 more |
| Apr 21, 2015 | Voluntary Action Indicated (VAI) | No citations listed |
| Mar 29, 2012 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Quality audits - Lack of or inadequate procedures (21 CFR 820.22) Lack of or inadequate procedures (21 CFR 820.100(a)) Design validation - user needs and intended uses (21 CFR 820.30(g)) and 3 more |
| Nov 7, 2008 | Voluntary Action Indicated (VAI) | Design validation - user needs and intended uses (21 CFR 820.30(g)) CAPA verification/validation of corrective/preventive action (21 CFR 820.100(a)(4)) Evaluation of suppliers, contractors, etc., requirements (21 CFR 820.50(a)) Lack of or inadequate process validation (21 CFR 820.75(a)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Sep 17, 2020 | Class II | Z-0310-2021 | A Li-Gator Single Use Laparoscopic Clip Applier, 10 mm X 29 cm - Product Usage: intended… Device malfunction causing the applier to jam, and clip loading failures. |
| Aug 25, 2020 | Class II | Z-0094-2021 | Aquas PowerFlow Tube Set - Product Usage: The device delivers sterile irrigation fluids to… The product design incorporating a motor driven impeller pump powered by one 9V alkaline battery inside the hand-piece did not receive… |
| Aug 21, 2018 | Class II | Z-2266-2020 | Genicon GENILook Model 200-005-151, 5mm diameter, 100mm length (one cannula, threaded)… and 13 more products There is potential for the plastic optical tip to become detached from the metal shaft. |
| Jun 27, 2018 | Class II | Z-2005-2020 | Genicon EZee Single Use Specimen Retrieval Bag - Product Usage: This device is disposable… Genicon is providing updated IFU labeling to users to provide clarity on how to use the device for multiple specimen retrieval. |
| Jul 19, 2017 | Class II | Z-1076-2018 | GENICON EZEE Retrieval, sterile, Rx only. Plastic handle is breaking off of shaft inhibiting proper removal per Instructions for Use. |
FDA record score
8 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
10 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Jan 17, 2020 | 6 x 0.25 | 1.5 |
| Inspection classified VAI | Oct 20, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-0310-2021, this FEI as the recalling firm | Sep 17, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0094-2021, this FEI as the recalling firm | Aug 25, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2266-2020, this FEI as the recalling firm | Aug 21, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-2005-2020, this FEI as the recalling firm | Jun 27, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1076-2018, this FEI as the recalling firm | Jul 19, 2017 | 4 x 0.25 | 1 |
| Inspection classified VAI | Apr 21, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Mar 29, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Nov 7, 2008 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Genicon, Inc. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Genicon, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Facility details
- Company
- Genicon, Inc.
- Location
- Winter Park, FL (US)
- Source
- inspections