FDA Inspection 817150 — SpineNet LLC.

SpineNet LLC. — FEI 3006755626

The FDA completed an inspection of SpineNet LLC. on February 1, 2013 in Maitland, FL. The inspection was classified OAI (Official Action Indicated — objectionable conditions were found and regulatory action is recommended). 14 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
February 1, 2013
Fiscal Year
2013
Product Type
Devices
Project Area
Compliance: Devices
Location
Maitland, FL (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
1472221 CFR 820.40Procedures not adequately established or maintained
232721 CFR 820.22Quality audits - Lack of or inadequate procedures
255721 CFR 820.30(c)Design input - documentation
265021 CFR 820.30(f)Design verification - Lack of or inadequate procedures
310221 CFR 820.30(h)Design transfer - Lack of or inadequate procedures
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
312121 CFR 820.80(b)Lack of or inadequate receiving acceptance procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
313921 CFR 820.140Lack of or inadequate procedures for handling
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
54621 CFR 820.75(a)Lack of or inadequate process validation
63021 CFR 803.17Lack of Written MDR Procedures