SpineNet LLC FDA Inspection Record
SpineNet LLC is an FDA-registered facility (FEI 3006755626) located in Maitland, FL, US. RegDataLab tracks 5 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Apr 19, 2024: No Action Indicated (NAI)
- Inspections on record
- 5 (1 OAI, 2 VAI, 2 NAI)
- Last OAI inspection
- Feb 1, 2013
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest May 31, 2013. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 2, latest Feb 5, 2013
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Apr 19, 2024 | No Action Indicated (NAI) | No citations listed |
| Oct 9, 2019 | Voluntary Action Indicated (VAI) | Design input - Lack of or inadequate procedures (21 CFR 820.30(c)) Quality audits - Lack of or inadequate procedures (21 CFR 820.22) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) |
| Oct 4, 2016 | No Action Indicated (NAI) | No citations listed |
| Feb 28, 2014 | Voluntary Action Indicated (VAI) | Lack of or inadequate process validation (21 CFR 820.75(a)) |
| Feb 1, 2013 | Official Action Indicated (OAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Procedures not adequately established or maintained (21 CFR 820.40) Quality audits - Lack of or inadequate procedures (21 CFR 820.22) Design input - documentation (21 CFR 820.30(c)) and 10 more |
Warning letters
- May 31, 2013: Warning letter listed on FDA's Data Dashboard. View compliance record (394073).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Feb 5, 2013 | Class II | Z-2103-2013 | SpineNet Anterior Cervical Cage, Lordotic, Peek (originally submitted as Daytona Anterior… The recall decision has been made because because it has recently been determined that the instructions for use (IFU) had certain statements… |
| Feb 4, 2013 | Class II | Z-2107-2013 | SpineNet Bone Marrow Aspiration Needle Kit (which is the Ranfac Bone Marrow Aspiration… SpineNet was marketing this device to be used with SpineNet ACC device without FDA clearance or approval. |
FDA record score
2 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
6 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Oct 9, 2019 | 6 x 0.25 | 1.5 |
| Inspection classified VAI | Feb 28, 2014 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | May 31, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2103-2013, this FEI as the recalling firm | Feb 5, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2107-2013, this FEI as the recalling firm | Feb 4, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Feb 1, 2013 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 28, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 28, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 28, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 28, 2026 | Newest record Nov 24, 2026 | Current |
| Recalls (Enforcement Report) | Nov 28, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing SpineNet LLC as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for SpineNet LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Facility details
- Company
- SpineNet LLC.
- Location
- Maitland, FL (US)
- Source
- device