FDA Inspection 817573 — VAX-D Medical Technologies LLC
The FDA completed an inspection of VAX-D Medical Technologies LLC on Feb 5, 2013 in Palm Harbor, FL. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 10 citations were issued during this inspection. Product area: Devices.
Inspection Details
- Classification
- Voluntary Action Indicated (VAI)
- End Date
- Feb 5, 2013
- Fiscal Year
- 2013
- Product Type
- Devices
- Project Area
- Postmarket Assurance: Devices
- Location
- Palm Harbor, FL (United States)
Citations
| ID | CFR | Description |
|---|---|---|
| 14713 | 21 CFR 820.198(a) | Lack of or inadequate complaint procedures |
| 14722 | 21 CFR 820.40 | Procedures not adequately established or maintained |
| 3130 | 21 CFR 820.100(a) | Lack of or inadequate procedures |
| 3149 | 21 CFR 820.180 | Availability |
| 3159 | 21 CFR 820.184 | DHR content |
| 3331 | 21 CFR 820.181 | DMR - not or inadequately maintained |
| 3677 | 21 CFR 820.30(g) | Design validation - software validation not performed |
| 4210 | 21 CFR 806.20(b)(2) | Device model identification |
| 479 | 21 CFR 820.50 | Purchasing controls, Lack of or inadequate procedures |
| 635 | 21 CFR 803.17(b)(1) | Info evaluated to determine if event was reportable |