FDA Inspection 821337 — Protech Professional Products, Inc.

Protech Professional Products, Inc. — FEI 1000137415

The FDA completed an inspection of Protech Professional Products, Inc. on Feb 27, 2013 in Boynton Beach, FL. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 11 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Voluntary Action Indicated (VAI)
End Date
Feb 27, 2013
Fiscal Year
2013
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Boynton Beach, FL (United States)

Citations

IDCFRDescription
1471221 CFR 820.184DHR - not or inadequately maintained
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
1472221 CFR 820.40Procedures not adequately established or maintained
237121 CFR 820.30(a)Design control - no procedures
312021 CFR 820.80(a)Lack of or inadequate procedures - Acceptance activities
313021 CFR 820.100(a)Lack of or inadequate procedures
313221 CFR 820.120Lack of or inadequate procedures for labeling
367121 CFR 820.25(a)Personnel
368021 CFR 820.70(a)Process control procedures, Lack of or inadequate procedures
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
63021 CFR 803.17Lack of Written MDR Procedures