FDA Inspection 864927 — Heartware, Inc.
The FDA completed an inspection of Heartware, Inc. on Jan 24, 2014 in Miami Lakes, FL. The inspection was classified OAI (Official Action Indicated — regulatory and/or administrative actions are recommended). 7 citations were issued during this inspection. Product area: Devices.
Inspection details
- Classification
- Official Action Indicated (OAI)
- End Date
- Jan 24, 2014
- Fiscal Year
- 2014
- Product Type
- Devices
- Project Area
- Compliance: Devices
- Location
- Miami Lakes, FL (United States)
Citations
| ID | CFR | Description |
|---|---|---|
| 3101 | 21 CFR 820.30(g) | Design validation- Lack of or inadequate procedures |
| 3102 | 21 CFR 820.30(h) | Design transfer - Lack of or inadequate procedures |
| 3117 | 21 CFR 820.70(i) | Software validation for automated processes |
| 3130 | 21 CFR 820.100(a) | Lack of or inadequate procedures |
| 3375 | 21 CFR 820.198(e) | Records of complaint investigation |
| 479 | 21 CFR 820.50 | Purchasing controls, Lack of or inadequate procedures |
| 546 | 21 CFR 820.75(a) | Lack of or inadequate process validation |