FDA Inspection 864927 — Heartware, Inc.

Heartware, Inc. — FEI 3007042319

The FDA completed an inspection of Heartware, Inc. on Jan 24, 2014 in Miami Lakes, FL. The inspection was classified OAI (Official Action Indicated — regulatory and/or administrative actions are recommended). 7 citations were issued during this inspection. Product area: Devices.

Inspection details

Classification
Official Action Indicated (OAI)
End Date
Jan 24, 2014
Fiscal Year
2014
Product Type
Devices
Project Area
Compliance: Devices
Location
Miami Lakes, FL (United States)

Citations

IDCFRDescription
310121 CFR 820.30(g)Design validation- Lack of or inadequate procedures
310221 CFR 820.30(h)Design transfer - Lack of or inadequate procedures
311721 CFR 820.70(i)Software validation for automated processes
313021 CFR 820.100(a)Lack of or inadequate procedures
337521 CFR 820.198(e)Records of complaint investigation
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
54621 CFR 820.75(a)Lack of or inadequate process validation