HeartWare, Inc. FDA Inspection Record
HeartWare, Inc. is an FDA-registered facility (FEI 3007042319) located in Miami Lakes, FL, US. RegDataLab tracks 5 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Jul 12, 2018: Voluntary Action Indicated (VAI)
- Inspections on record
- 5 (2 OAI, 2 VAI, 1 NAI)
- Last OAI inspection
- May 4, 2016
- Warning letters
- 1 listed by FDA for this FEI, latest Jun 2, 2014
- Recalls as the recalling firm
- 35, latest Oct 31, 2023
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jul 12, 2018 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Training - Lack of or inadequate procedures (21 CFR 820.25(b)) Design control - no procedures (21 CFR 820.30(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) and 3 more |
| May 4, 2016 | Official Action Indicated (OAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design transfer - Lack of or inadequate procedures (21 CFR 820.30(h)) Lack of or inadequate procedures (21 CFR 820.100(a)) Servicing - Lack of or inadequate procedures (21 CFR 820.200(a)) and 3 more |
| Jan 24, 2014 | Official Action Indicated (OAI) | Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Design transfer - Lack of or inadequate procedures (21 CFR 820.30(h)) Software validation for automated processes (21 CFR 820.70(i)) Lack of or inadequate procedures (21 CFR 820.100(a)) and 3 more |
| Aug 30, 2013 | No Action Indicated (NAI) | No citations listed |
| Jul 20, 2011 | Voluntary Action Indicated (VAI) | Quality audits - Lack of or inadequate procedures (21 CFR 820.22) Training - Lack of or inadequate procedures (21 CFR 820.25(b)) Lack of or inadequate procedures (21 CFR 820.100(a)) Procedures for product review,disposition lack of/inadequate (21 CFR 820.90(b)(1)) and 1 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Oct 31, 2023 | Class II | Z-0475-2024 | HVAD (HeartWare Ventricular Assist Device) STERILE IMPLANT KIT, Model Number 1100 and 8 more products Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the… |
| May 16, 2023 | Class II | Z-1993-2023 | HeartWare Ventricular Assist Device System.
a) Monitor 1521US Explore Tech,
b) Monitor… Medtronic has identified that the logfiles downloaded from the recently updated Model 1521 monitors (Serial Numbers: MON4xxxxxx for OUS and… |
| Nov 29, 2022 | Class II | Z-0742-2023 | HeartWare Model 1100 and 8 more products Over time, the driveline cover may become hardened, creating difficulty or an inability to slide the cover back to access the driveline to… |
| Jun 28, 2022 | Class I | Z-1512-2022 | HeartWare Ventricular Assist Device (HVAD) System Batteries, Product Number 1650DE Battery performance issues. Battery electrical faults render it unable to power the controller, unable to accept charge from the battery… |
| Mar 30, 2022 | Class II | Z-1102-2022 | HeartWare Ventricular Assist Device (HVAD) System, Pump Implant Kit, Model Number1104 and 5 more products Finnish and Turkish translation errors in HVAD controller and monitor displays as will as Instructions for Use, Patient Manual, and… |
| Jan 27, 2022 | Class II | Z-0726-2022 | HeartWare Ventricular Assist Device (HVAD) System Updated cleaning instructions are being provided for the Controller AC Adapter, DC Adapter, and Battery to provide clarity to avoid cleaning… |
| Aug 6, 2021 | Class II | Z-2478-2021 | Medtronic HVAD Pump Implant Kits There is more force required to slide the driveline cover (boot cover) away from the connector after completion of a Driveline Strain Relief… |
| Jun 3, 2021 | Class I | Z-2188-2021 | HeartWare HVAD Pump Implant Kit, REF 1103 and 9 more products Medtronic is stopping the distribution and sale of the HVAD System due to a growing body of observational clinical comparisons that shows a… |
| May 28, 2021 | Class II | Z-2072-2021 | HeartWare, HVAD Pump Implant Kit, REF MCS1705PU - Product Usage: indicated for hemodynamic… The pump has an impeller with a shroud height that did not meet the lower control limit. |
| Feb 26, 2021 | Class I | Z-1336-2021 | Medtronic HVAD Controller, Product Numbers: 1400, 1401, 1401AU, 1401BR, 1401CA, 1401CH… and 5 more products There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor… |
FDA record score
63 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
40 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class I recall Z-1512-2022, this FEI as the recalling firm | Jun 28, 2022 | 15 x 0.6 | 9 |
| Class I recall Z-2188-2021, this FEI as the recalling firm | Jun 3, 2021 | 15 x 0.25 | 3.8 |
| Class I recall Z-1336-2021, this FEI as the recalling firm | Feb 26, 2021 | 15 x 0.25 | 3.8 |
| Class I recall Z-1426-2021, this FEI as the recalling firm | Feb 26, 2021 | 15 x 0.25 | 3.8 |
| Class I recall Z-0946-2021, this FEI as the recalling firm | Nov 19, 2020 | 15 x 0.25 | 3.8 |
| Class I recall Z-2055-2020, this FEI as the recalling firm | Apr 3, 2020 | 15 x 0.25 | 3.8 |
| Class I recall Z-1427-2020, this FEI as the recalling firm | Jan 23, 2020 | 15 x 0.25 | 3.8 |
| Class I recall Z-1903-2018, this FEI as the recalling firm | May 2, 2018 | 15 x 0.25 | 3.8 |
| Class I recall Z-1832-2017, this FEI as the recalling firm | Mar 10, 2017 | 15 x 0.25 | 3.8 |
| Class I recall Z-1538-2017, this FEI as the recalling firm | Feb 3, 2017 | 15 x 0.25 | 3.8 |
| Class II recall Z-0475-2024, this FEI as the recalling firm | Oct 31, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1993-2023, this FEI as the recalling firm | May 16, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-0742-2023, this FEI as the recalling firm | Nov 29, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-1102-2022, this FEI as the recalling firm | Mar 30, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-0726-2022, this FEI as the recalling firm | Jan 27, 2022 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Jul 12, 2018 | 6 x 0.25 | 1.5 |
| Class II recall Z-2478-2021, this FEI as the recalling firm | Aug 6, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-2072-2021, this FEI as the recalling firm | May 28, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-1438-2021, this FEI as the recalling firm | Feb 26, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-1469-2021, this FEI as the recalling firm | Feb 26, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0582-2019, this FEI as the recalling firm | Nov 19, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0067-2019, this FEI as the recalling firm | Aug 28, 2018 | 4 x 0.25 | 1 |
| Class I recall Z-0006-2017, this FEI as the recalling firm | Aug 17, 2016 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-0005-2017, this FEI as the recalling firm | Jun 8, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | May 4, 2016 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-1887-2016, this FEI as the recalling firm | Jan 7, 2016 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-1604-2015, this FEI as the recalling firm | Apr 29, 2015 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-1698-2015, this FEI as the recalling firm | Apr 29, 2015 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-1726-2015, this FEI as the recalling firm | Apr 29, 2015 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-1731-2015, this FEI as the recalling firm | Apr 29, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1751-2015, this FEI as the recalling firm | Apr 29, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1917-2015, this FEI as the recalling firm | Apr 29, 2015 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-1131-2015, this FEI as the recalling firm | Jan 16, 2015 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Jun 2, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1607-2014, this FEI as the recalling firm | Apr 16, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Jan 24, 2014 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-1471-2014, this FEI as the recalling firm | Dec 6, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1323-2013, this FEI as the recalling firm | Dec 27, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jul 20, 2011 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1813-2013, this FEI as the recalling firm | Date not listed | No date listed, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 23, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 23, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 23, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 23, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 23, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing HeartWare, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check HeartWare, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Heartware, Inc.
- Location
- Miami Lakes, FL (US)
- Source
- device