898177

Natus Medical Incorporated — FEI 3018859

Inspection Details

Classification
No Action Indicated (NAI) (NAI)
End Date
September 10, 2014
Fiscal Year
2014
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Seattle, WA (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
1471621 CFR 820.30(f)Design verification - output does not meet input requirement
310121 CFR 820.30(g)Design validation- Lack of or inadequate procedures
310421 CFR 820.30(j)Design history file
313021 CFR 820.100(a)Lack of or inadequate procedures
333121 CFR 820.181DMR - not or inadequately maintained
369621 CFR 820.100(b)Documentation