Natus Medical Incorporated FDA Inspection Record
Natus Medical Incorporated is an FDA-inspected facility (FEI 3018859) located in Seattle, WA, US. RegDataLab tracks 9 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Aug 13, 2018: Voluntary Action Indicated (VAI)
- Inspections on record
- 9 (1 OAI, 3 VAI, 5 NAI)
- Last OAI inspection
- Sep 10, 2014
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Apr 10, 2015. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 13, latest May 22, 2017
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Aug 13, 2018 | Voluntary Action Indicated (VAI) | Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
| Jun 27, 2017 | No Action Indicated (NAI) | No citations listed |
| Jul 27, 2016 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design input - documentation (21 CFR 820.30(c)) Design verification - Lack of or inadequate procedures (21 CFR 820.30(f)) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) and 8 more |
| Sep 10, 2014 | Official Action Indicated (OAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design verification - output does not meet input requirement (21 CFR 820.30(f)) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Design history file (21 CFR 820.30(j)) and 3 more |
| Jul 11, 2012 | No Action Indicated (NAI) | No citations listed |
| Feb 7, 2012 | No Action Indicated (NAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Design validation - simulated testing (21 CFR 820.30(g)) Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) |
| Sep 2, 2010 | No Action Indicated (NAI) | No citations listed |
| Jun 9, 2009 | No Action Indicated (NAI) | No citations listed |
| Oct 8, 2008 | Voluntary Action Indicated (VAI) | Acceptance records (21 CFR 820.184(d)) Calibration documentation (21 CFR 820.72(b)(2)) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) Design verification - documentation (21 CFR 820.30(f)) and 1 more |
Warning letters
- Apr 10, 2015: Warning letter listed on FDA's Data Dashboard. View compliance record (442835).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| May 22, 2017 | Class II | Z-2701-2017 | The Olympic Pasteurmatic 3000 and Olympic Pasteurmatic 3500 are intended for cleaning and… Olympic Pasteurmatic 3000/3500 Systems are recalled because they do not fully meet todays regulatory requirements. Given the age of the… |
| Apr 3, 2017 | Class II | Z-1924-2017 | Second generation neoBLUE LED Phototherapy Light (neoBLUE 2) sold between 2003 and 2008… The second generation neoBLUE 2 lights with a black band, sold from 2003 to 2008, are recalled so it will eliminate the possibility that a… |
| Oct 28, 2016 | Class II | Z-0770-2017 | Natus neoBLUE blanket LED Phototherapy system, neonatal phototherapy. Used to provide… neoBLUE blanket LED Phototherapy System distributed since release of the product in 2011 is recalled due to customer complaints regarding… |
| Sep 16, 2016 | Class II | Z-0769-2017 | Updated Natus neoBLUE blanket LED Phototherapy Systems. It provides phototherapy… The firm sent the Technical Bulletin to consignees who received the reliability updated neoBLUE blanket LED Phototherapy Systems between… |
| Aug 25, 2016 | Class II | Z-0056-2017 | The Medix NatalCare LX Intensive Care Incubator. Neonatal incubator.
Item numbers… NatalCare LX incubators may have a defective base where the weld holding the wheel bushing to the tube is inadequate. There were several… |
| Apr 22, 2016 | Class II | Z-1878-2016 | Replacement HEPA filter for Natus Sterile-Drier Model 43 and Model 44. Filter Catalogue… The replacement HEPA Filters have been manufactured with sealant that may not be properly cured. Improperly cured sealant can emit a toxic… |
| Apr 11, 2016 | Class II | Z-2001-2016 | Natus neoBLUE Phototherapy system, Catalogue Part Number: 010066.
Top Assembly Part… The firm is adding the warning statement to the Instruction for Use for the neoBLUE LED Phototherapy System and provide a laminated Quick… |
| Jan 8, 2016 | Class II | Z-1363-2016 | Medix NatalCare ST-LX Intensive Care Incubator. Neonatal incubator. NatalCare LX Incubators were manufactured using a defective oval port hinge assembly with the result that the screws holding the port in… |
| Aug 5, 2015 | Class II | Z-0149-2016 | Natus neoBLUE 3 Phototherapy system or Neonatal Phototherapy, Catalogue Part Number… The neoBLUE Instruction For Use (IFU) and Service Manuals for the neoBLUE 3 were revised to show the cleared intensity statement settings so… |
| Mar 16, 2015 | Class II | Z-1412-2015 | Natus neoBLUE blanket LED Phototherapy system which includes the following… neoBLUE blanket LED Phototherapy System is recalled because Natus has received feedback from some customers about early failure of the… |
FDA record score
4 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
18 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Aug 13, 2018 | 6 x 0.25 | 1.5 |
| Class II recall Z-2701-2017, this FEI as the recalling firm | May 22, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1924-2017, this FEI as the recalling firm | Apr 3, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0770-2017, this FEI as the recalling firm | Oct 28, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0769-2017, this FEI as the recalling firm | Sep 16, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0056-2017, this FEI as the recalling firm | Aug 25, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jul 27, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1878-2016, this FEI as the recalling firm | Apr 22, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2001-2016, this FEI as the recalling firm | Apr 11, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1363-2016, this FEI as the recalling firm | Jan 8, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0149-2016, this FEI as the recalling firm | Aug 5, 2015 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Apr 10, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1412-2015, this FEI as the recalling firm | Mar 16, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0847-2015, this FEI as the recalling firm | Nov 19, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Sep 10, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1843-2012, this FEI as the recalling firm | May 21, 2012 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-0447-2013, this FEI as the recalling firm | May 9, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Oct 8, 2008 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Natus Medical Incorporated as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Natus Medical Incorporated yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Natus Medical Incorporated
- Location
- Seattle, WA (US)
- Source
- inspections