FDA Inspection 920985 — Smith & Nephew Inc.

Smith & Nephew Inc. — FEI 3003604053

The FDA completed an inspection of Smith & Nephew Inc. on March 26, 2015 in Andover, MA. The inspection was classified OAI (Official Action Indicated — objectionable conditions were found and regulatory action is recommended). 8 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Official Action Indicated (OAI) (OAI)
End Date
March 26, 2015
Fiscal Year
2015
Product Type
Devices
Project Area
Postmarket Assurance: Devices
Location
Andover, MA (United States)

Citations

IDCFRDescription
232721 CFR 820.22Quality audits - Lack of or inadequate procedures
265021 CFR 820.30(f)Design verification - Lack of or inadequate procedures
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
328221 CFR 820.90(a)Nonconforming product, Lack of or inadequate procedures
366621 CFR 820.20(c)Management review - Lack of or inadequate procedures
367821 CFR 820.30(g)Design Validation - Risk analysis not performed/inadequate
73221 CFR 803.50(a)(2)Individual Report of Malfunction