SMITH & NEPHEW INC. FDA Inspection Record
SMITH & NEPHEW INC. is an FDA-registered facility (FEI 3003604053) located in ANDOVER, MA, US. RegDataLab tracks 4 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Oct 26, 2021: No Action Indicated (NAI)
- Inspections on record
- 4 (1 OAI, 1 VAI, 2 NAI)
- Last OAI inspection
- Mar 26, 2015
- Warning letters
- 1 listed by FDA for this FEI, latest Apr 30, 2015
- Recalls as the recalling firm
- 13, latest Oct 29, 2024
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Oct 26, 2021 | No Action Indicated (NAI) | No citations listed |
| Nov 9, 2016 | No Action Indicated (NAI) | No citations listed |
| Mar 26, 2015 | Official Action Indicated (OAI) | Quality audits - Lack of or inadequate procedures (21 CFR 820.22) Design verification - Lack of or inadequate procedures (21 CFR 820.30(f)) Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Lack of or inadequate procedures (21 CFR 820.100(a)) and 4 more |
| Aug 6, 2009 | Voluntary Action Indicated (VAI) | Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Documentation - review in response to changes or deviations (21 CFR 820.75(c)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Oct 29, 2024 | Class II | Z-0581-2025 | BIORAPTOR Suture Anchors. Tendon/ligament, non-bioabsorbable bone anchor. Sterile barrier breach due to inadequate packaging design that could not hold the geometry of the device. |
| Aug 12, 2024 | Class II | Z-3096-2024 | ROTATION MEDICAL TENDON STAPLES (8) (broad indication)-Intended for fixation of prosthetic… and 1 more product Product packaging process may result in an improper or incomplete seal of the outer foil pouch surrounding the inner Tyvek pouch… |
| Apr 8, 2021 | Class II | Z-1510-2021 | CANNU-Flex SILK Interference Screws 6 mmx 30 mm - Product Usage: used for interference… Mislabeled box contained a 6mm x 20mm screw instead of 6mm x 30mm screw |
| Sep 21, 2020 | Class II | Z-0301-2021 | 5.5MM Arthrogarde Hip Access Cannula The J Lock feature on the cannula body was manufactured incorrectly, preventing the cannula from locking into an irrigation extender or… |
| Sep 9, 2020 | Class II | Z-0136-2021 | TWINFIX 5.0MM AB SUTURE ANCHOR WITH TWO
Absorbable Suture anchors
Product Number… and 39 more products Product packaging improper or incomplete seal of the pouch surrounding the sterile product may result in a breach of the sterile barrier… |
| Jun 24, 2020 | Class II | Z-2702-2020 | Smith & Nephew T-F1X0 RCG Drill Pac, 2 mm Drill & (2) Sleeves - Product Usage: indicated… The T-Fix sterile package contains only one sleeve instead of two, The device will not perform as intended for the second bone tunnel due… |
| Jun 24, 2020 | Class II | Z-2746-2020 | SUTUREFIX ULTRA Suture, Product number 72203855 A packaging error resulted in a XL drill being presented for use instead of the S drill. The XL drill is one inch longer than the S drill. |
| Jan 11, 2016 | Class II | Z-0793-2016 | Smith & Nephew BEAVER Blade 4.0mm
Part Number: 72203307
Orthopedic surgical instrument… Blade exhibiting fracture during hip arthroscopy procedures |
| Apr 30, 2015 | Class II | Z-1822-2015 | ULTRA-FAST FIX Knot Pusher Suture Cutters
Product Number: 72201537
Intended for… Sterility of device maybe compromised |
| Jun 2, 2014 | Class II | Z-2045-2014 | SUTUREFIX Ultra Anchor XL with one (#2) Ultrabraid Blue
Product Code… and 4 more products Sterility of device is compromised due to breach in sterile barrier |
FDA record score
10 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
16 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0581-2025, this FEI as the recalling firm | Oct 29, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-3096-2024, this FEI as the recalling firm | Aug 12, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-1510-2021, this FEI as the recalling firm | Apr 8, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0301-2021, this FEI as the recalling firm | Sep 21, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0136-2021, this FEI as the recalling firm | Sep 9, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2702-2020, this FEI as the recalling firm | Jun 24, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2746-2020, this FEI as the recalling firm | Jun 24, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0793-2016, this FEI as the recalling firm | Jan 11, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1822-2015, this FEI as the recalling firm | Apr 30, 2015 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Apr 30, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Mar 26, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2045-2014, this FEI as the recalling firm | Jun 2, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2215-2013, this FEI as the recalling firm | Aug 6, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2144-2012, this FEI as the recalling firm | Jul 14, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2165-2012, this FEI as the recalling firm | Jul 3, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Aug 6, 2009 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing SMITH & NEPHEW INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check SMITH & NEPHEW INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Smith & Nephew Inc.
- Location
- ANDOVER, MA (US)
- Source
- device