SMITH & NEPHEW INC. FDA Inspection Record

FEI 3003604053 · Inspections, Form 483s, warning letters and import refusals

SMITH & NEPHEW INC. is an FDA-registered facility (FEI 3003604053) located in ANDOVER, MA, US. RegDataLab tracks 4 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Oct 26, 2021: No Action Indicated (NAI)
Inspections on record
4 (1 OAI, 1 VAI, 2 NAI)
Last OAI inspection
Mar 26, 2015
Warning letters
1 listed by FDA for this FEI, latest Apr 30, 2015
Recalls as the recalling firm
13, latest Oct 29, 2024

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Oct 26, 2021No Action Indicated (NAI)No citations listed
Nov 9, 2016No Action Indicated (NAI)No citations listed
Mar 26, 2015Official Action Indicated (OAI)Quality audits - Lack of or inadequate procedures (21 CFR 820.22)
Design verification - Lack of or inadequate procedures (21 CFR 820.30(f))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Lack of or inadequate procedures (21 CFR 820.100(a))
and 4 more
Aug 6, 2009Voluntary Action Indicated (VAI)Design validation- Lack of or inadequate procedures (21 CFR 820.30(g))
Documentation - review in response to changes or deviations (21 CFR 820.75(c))

Recalls

Showing the 10 most recent of 13 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Oct 29, 2024Class IIZ-0581-2025BIORAPTOR Suture Anchors. Tendon/ligament, non-bioabsorbable bone anchor.
Sterile barrier breach due to inadequate packaging design that could not hold the geometry of the device.
Aug 12, 2024Class IIZ-3096-2024ROTATION MEDICAL TENDON STAPLES (8) (broad indication)-Intended for fixation of prosthetic… and 1 more product
Product packaging process may result in an improper or incomplete seal of the outer foil pouch surrounding the inner Tyvek pouch…
Apr 8, 2021Class IIZ-1510-2021CANNU-Flex SILK Interference Screws 6 mmx 30 mm - Product Usage: used for interference…
Mislabeled box contained a 6mm x 20mm screw instead of 6mm x 30mm screw
Sep 21, 2020Class IIZ-0301-20215.5MM Arthrogarde Hip Access Cannula
The J Lock feature on the cannula body was manufactured incorrectly, preventing the cannula from locking into an irrigation extender or…
Sep 9, 2020Class IIZ-0136-2021TWINFIX 5.0MM AB SUTURE ANCHOR WITH TWO Absorbable Suture anchors Product Number… and 39 more products
Product packaging improper or incomplete seal of the pouch surrounding the sterile product may result in a breach of the sterile barrier…
Jun 24, 2020Class IIZ-2702-2020Smith & Nephew T-F1X0 RCG Drill Pac, 2 mm Drill & (2) Sleeves - Product Usage: indicated…
The T-Fix sterile package contains only one sleeve instead of two, The device will not perform as intended for the second bone tunnel due…
Jun 24, 2020Class IIZ-2746-2020SUTUREFIX ULTRA Suture, Product number 72203855
A packaging error resulted in a XL drill being presented for use instead of the S drill. The XL drill is one inch longer than the S drill.
Jan 11, 2016Class IIZ-0793-2016Smith & Nephew BEAVER Blade 4.0mm Part Number: 72203307 Orthopedic surgical instrument…
Blade exhibiting fracture during hip arthroscopy procedures
Apr 30, 2015Class IIZ-1822-2015ULTRA-FAST FIX Knot Pusher Suture Cutters Product Number: 72201537 Intended for…
Sterility of device maybe compromised
Jun 2, 2014Class IIZ-2045-2014SUTUREFIX Ultra Anchor XL with one (#2) Ultrabraid Blue Product Code… and 4 more products
Sterility of device is compromised due to breach in sterile barrier

FDA record score

10 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 23, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

16 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-0581-2025, this FEI as the recalling firmOct 29, 20244 x 0.62.4
Class II recall Z-3096-2024, this FEI as the recalling firmAug 12, 20244 x 0.62.4
Class II recall Z-1510-2021, this FEI as the recalling firmApr 8, 20214 x 0.251
Class II recall Z-0301-2021, this FEI as the recalling firmSep 21, 20204 x 0.251
Class II recall Z-0136-2021, this FEI as the recalling firmSep 9, 20204 x 0.251
Class II recall Z-2702-2020, this FEI as the recalling firmJun 24, 20204 x 0.251
Class II recall Z-2746-2020, this FEI as the recalling firmJun 24, 20204 x 0.251
Class II recall Z-0793-2016, this FEI as the recalling firmJan 11, 201610 years or older, so it adds no points0
Class II recall Z-1822-2015, this FEI as the recalling firmApr 30, 201510 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsApr 30, 201510 years or older, so it adds no points0
Inspection classified OAIMar 26, 201510 years or older, so it adds no points0
Class II recall Z-2045-2014, this FEI as the recalling firmJun 2, 201410 years or older, so it adds no points0
Class II recall Z-2215-2013, this FEI as the recalling firmAug 6, 201310 years or older, so it adds no points0
Class II recall Z-2144-2012, this FEI as the recalling firmJul 14, 201210 years or older, so it adds no points0
Class II recall Z-2165-2012, this FEI as the recalling firmJul 3, 201210 years or older, so it adds no points0
Inspection classified VAIAug 6, 200910 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 22, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 22, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 22, 2026Newest record Nov 13, 2026Current
Warning letter textNov 22, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 22, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing SMITH & NEPHEW INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check SMITH & NEPHEW INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Smith & Nephew Inc.
Location
ANDOVER, MA (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.