FDA Inspection 949804 — Practica Medical Manufacturing, Inc

Practica Medical Manufacturing, Inc — FEI 1000119184

The FDA completed an inspection of Practica Medical Manufacturing, Inc on Nov 19, 2015 in Stuart, FL. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions or practices were found, but the agency is not prepared to take or recommend any administrative or regulatory action). 5 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Voluntary Action Indicated (VAI)
End Date
Nov 19, 2015
Fiscal Year
2016
Product Type
Devices
Project Area
Compliance: Devices
Location
Stuart, FL (United States)

Citations

IDCFRDescription
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
1471821 CFR 820.30(g)Design validation - Risk analysis
232721 CFR 820.22Quality audits - Lack of or inadequate procedures
47921 CFR 820.50Purchasing controls, Lack of or inadequate procedures
54621 CFR 820.75(a)Lack of or inadequate process validation