Z-0078-2023 Class II Ongoing

Recalled by Olympus Corporation Of The Americas — Center Valley, PA

FDA device recall Z-0078-2023 was initiated by Olympus Corporation Of The Americas on Sep 16, 2022 and is designated Class II. Reason for recall: Update to the Instruction for Use (IFU): Olympus is adding a new warning against the use of accessories which are not listed in the IFU. There have been reported adverse events involving the use of… The recall status is ongoing. Affected quantity: 4184 US; 191 OUS.

Recall Details

Product Type
Devices
Report Date
Oct 19, 2022
Initiation Date
Sep 16, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4184 US; 191 OUS

Product Description

Berkeley VC-10 Vacuum Curettage System, Model No. VC-10. For transcervical aspiration of the uterine cavity.

Reason for Recall

Update to the Instruction for Use (IFU): Olympus is adding a new warning against the use of accessories which are not listed in the IFU. There have been reported adverse events involving the use of collection canisters which were not designed for use with the VC-10, which can lead to serious injuries as it does not fit properly with the VC-10.

Distribution Pattern

Domestic distribution nationwide. Foreign distribution to Canada and Japan.

Code Information

UDI-DI: 00821925010475 All units are included.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0078-2023
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 24, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.