Z-3247-2026 Class II Ongoing

Recalled by Olympus Corporation Of The Americas — Center Valley, PA

FDA device recall Z-3247-2026 was initiated by Olympus Corporation Of The Americas on Aug 19, 2026 and is designated Class II. Reason for recall: Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive. The recall status is ongoing. Affected quantity: 816 units.

Recall Details

Product Type
Devices
Report Date
Sep 30, 2026
Initiation Date
Aug 19, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
816 units

Product Description

Olympus BiCOAG Hemostasis Probe 7Fr Fixed Pin Connector. Model/Catalog Number: CD-B622LA.

Reason for Recall

Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, Latin America, Singapore, Hong Kong, and EU.

Code Information

Lot Code: Model: CD-B622LA. UDI-DI: 00821925039452. Lot Numbers: All unexpired lots