Z-0085-2021 Class II Ongoing

Recalled by Roche Diagnostics Operations, Inc. — Indianapolis, IN

Recall Details

Product Type
Devices
Report Date
October 14, 2020
Initiation Date
August 31, 2020
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2623 (US); 3735 (ex-US)

Product Description

BenchMark ULTRA Advanced Staining System (IVD)- Material Number: 05342716001

Reason for Recall

Potential for Fluid leaks within the instruments, creating an electrical short inside the power socket, which generates heat, smoke, and eventually burning and melting of the electrical cord connector and socket

Distribution Pattern

Nationwide Foreign: Canada, Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Latvia, Lithuania, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, UK, Algeria, Argentina, Aruba, Australia, Bahrain, Bosnia-Herz., Brazil, Chile, China, Colombia, Costa Rica, Cuba, Dominican Republic, Ecuador, Egypt, El Salvador, Hong Kong, India, Israel, Japan, Jordan, Kazakhstan, Kuwait, Malaysia, Martinque, Mexico, Monaco, Morocco, Namibia, New Caledonia, New Zealand, Oman, Pakistan, Panama, Peru, Philippines, Russia, Saudi Arabia, Singapore, S. Africa, S. Korea, Taiwan, Thailand, Ukraine, UAE, Vietnam

Code Information

All units