Z-0093-2025 Class III Ongoing

Recalled by Beckman Coulter, Inc. — Miami, FL

FDA device recall Z-0093-2025 was initiated by Beckman Coulter, Inc. on Jul 16, 2024 and is designated Class III. Reason for recall: Beckman Coulter has become aware of an increase in customer complaints for intermittent IRISpec CB negative glucose control failure, resulting in false positive control results when used in… The recall status is ongoing. Affected quantity: 5589 units.

Recall Details

Product Type
Devices
Report Date
Oct 23, 2024
Initiation Date
Jul 16, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5589 units

Product Description

Beckman Coulter IRISPEC CA/CB/CC Urine Chemistry Controls, Catalog Number REF 800-7211 and 800-7702, in vitro diagnostic device.

Reason for Recall

Beckman Coulter has become aware of an increase in customer complaints for intermittent IRISpec CB negative glucose control failure, resulting in false positive control results when used in conjunction with iChemVELOCITY Urine Chemistry Strips (PNs 800-7204 and 800-7212) on the iChemVELOCITY Analyzer. This issue may delay the reporting of patient results, with a worst-case scenario of a delay in diagnosis or treatment of metabolic disorders, kidney function abnormalities, urinary tract infections, and liver function.

Distribution Pattern

worldwide

Code Information

UDI/DI 10837461002611, Lot Numbers: 432-24 and higher

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0093-2025
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 24, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.