Z-0115-2023 Class II Ongoing

Recalled by Abbott Vascular — Santa Clara, CA

Recall Details

Product Type
Devices
Report Date
October 26, 2022
Initiation Date
September 8, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
18,042

Product Description

Product/Part: MitraClip XTR Clip Delivery System US/CDS0601-XTR, MitraClip G4 Clip Delivery System NT US/CDS0701-NT, MitraClip G4 Clip Delivery System NTW US/CDS0701-NTW, MitraClip G4 Clip Delivery System XT US/CDS0701-XT, MitraClip G4 Clip Delivery System XTW US/CDS0701-XTW. For cardiac procedures.

Reason for Recall

An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to recurrence of severe mitral regurgitation, device embolization, or tissue injury.

Distribution Pattern

Worldwide distribution - US Nationwide including in the states of PR, WA, NE, MO, IA, KS, TN, MN, ND, MD, VA, MA, PA, DC, LA, KY, CA, NV, IL, WI, TX, CT, VT, AL, AR, GA, MS, MI, OH, FL, NC, NJ, CO, NY, DE, OR, SC, IN, HI, AZ, NH, ME, WV, OK, ID, UT, MT, NM, RI, AK, SD and the countries of AE, AR, AT, AU, BE, BG, BR, BY, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GB, GR, HK, HU, ID, IE, IL, IN, IR, IT, JO, JP, KR, KW, LB, LU, MT, MX, MY, NL, NO, NZ, OM, PE, PH, PK, PL, PT, QA, RE, RO, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW.

Code Information

Part/UDI: CDS0601-XTR/08717648226366, CDS0701-NT/08717648230943, CDS0701-NTW/08717648230967, CDS0701-XT/08717648230981, CDS0701-XTW/08717648231001, All lots.