Z-0116-2023 Class II Ongoing

Recalled by Abbott Vascular — Santa Clara, CA

FDA device recall Z-0116-2023 was initiated by Abbott Vascular on Sep 8, 2022 and is designated Class II. Reason for recall: An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain… The recall status is ongoing. Affected quantity: 20,045.

Recall details

Product Type
Devices
Report Date
Oct 26, 2022
Initiation Date
Sep 8, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
20,045

Product description

Product/Part: MitraClip NTR Clip Delivery System US/CDS0601-NTR, MitraClip NTR Clip Delivery System CE/CDS0602-NTR, MitraClip XTR Clip Delivery System CE/CDS0602-XTR, MitraClip G4 Clip Delivery System NT CE/CDS0702-NT, MitraClip G4 Clip Delivery System NTW CE/CDS0702-NTW, MitraClip G4 Clip Delivery System XT CE/CDS0702-XT, MitraClip G4 Clip Delivery System XTW CE/CDS0702-XTW, MitraClip G4 Clip Delivery Sys NT ROW/CDS0705-NT, MitraClip G4 Clip Delivery Sys NTW ROW/CDS0705-NTW, MitraClip G4 Clip Delivery Sys XT ROW/CDS0705-XT, MitraClip G4 Clip Delivery Sys XTW ROW/CDS0705-XTW, TriClip XT Clip Delivery Sys CE/TCDS0202-XT, TriClip NT Clip Delivery Sys CE/TCDS0202-NT, TriClip XT Clip Delivery Sys ROW/TCDS0205-XT, TriClip NT Clip Delivery Sys ROW/TCDS0205-NT, TriClip G4 NT Delivery System CE/TCDS0302-NT, TriClip G4 NTW Delivery System CE/TCDS0302-NTW, TriClip G4 XT Delivery System CE/TCDS0302-XT, TriClip G4 XTW Delivery System CE/TCDS0302-XTW, TriClip G4 NT Delivery System ROW/TCDS0305-NT, TriClip G4 NTW Delivery System ROW/TCDS0305-NTW, TriClip G4 XT Delivery System ROW/TCDS0305-XT, TriClip G4 XTW Delivery System ROW/TCDS0305-XTW, TriClip NT Clip Delivery Sys CE/TCDS0202-NT, TriClip NT Clip Delivery Sys ROW/TCDS0205-NT. For cardiac procedures.

Reason for recall

An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to recurrence of severe mitral regurgitation, device embolization, or tissue injury.

Distribution pattern

Worldwide distribution - US Nationwide including in the states of PR, WA, NE, MO, IA, KS, TN, MN, ND, MD, VA, MA, PA, DC, LA, KY, CA, NV, IL, WI, TX, CT, VT, AL, AR, GA, MS, MI, OH, FL, NC, NJ, CO, NY, DE, OR, SC, IN, HI, AZ, NH, ME, WV, OK, ID, UT, MT, NM, RI, AK, SD and the countries of AE, AR, AT, AU, BE, BG, BR, BY, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GB, GR, HK, HU, ID, IE, IL, IN, IR, IT, JO, JP, KR, KW, LB, LU, MT, MX, MY, NL, NO, NZ, OM, PE, PH, PK, PL, PT, QA, RE, RO, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW.

Code information

Part/UDI-GTIN: CDS0601-NTR/08717648226342, CDS0602-NTR/08717648226359, CDS0602-XTR/08717648226373, CDS0702-NT/08717648230950, CDS0702-NTW/08717648230974, CDS0702-XT/08717648230998, CDS0702-XTW/08717648231018, CDS0705-NT/08717648288616, CDS0705-NTW/08717648288630, CDS0705-XT/08717648288623, CDS0705-XTW/08717648288647, TCDS0202-XT/08717648229978, TCDS0202-NT/08717648229985, TCDS0205-XT/08717648288708, TCDS0205-NT/08717648288692, TCDS0302-NT/08717648334238, TCDS0302-NTW/08717648334252, TCDS0302-XT/08717648334245, TCDS0302-XTW/08717648334269, TCDS0305-NT/08717648334313, TCDS0305-NTW/08717648334337, TCDS0305-XT/08717648334320, TCDS0305-XTW/08717648334344, TCDS0202-NT/08717648229985, TCDS0205-NT/08717648288692 All lots.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0116-2023
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0116-2023. Report a data issue: hello@regdatalab.com.