Z-0730-2025 Class II Ongoing
FDA device recall Z-0730-2025 was initiated by Abbott Vascular Inc on Nov 6, 2024 and is designated Class II. Reason for recall: Inflation Device used during cardiovascular procedures may leak in the indeflator at the hose rotating luer or at the gauge, which could lead to air embolism. The recall status is ongoing. Affected quantity: 1,418.
Recall details
- Product Type
- Devices
- Report Date
- Jan 1, 2025
- Initiation Date
- Nov 6, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1,418
Product description
20/30 INDEFLATOR, REF: 1000184, for use during cardiovascular procedures in conjunction with interventional devices such as balloon dilatation catheters to create and monitor pressure in the balloon.
Reason for recall
Inflation Device used during cardiovascular procedures may leak in the indeflator at the hose rotating luer or at the gauge, which could lead to air embolism.
Distribution pattern
US: IA, MD, DC, IN, CA, NV, TX, CT, GA, MS, NY, OH, FL, NC, MI, KY, PA, VA, IL, OK, TN, UT, WA, WI, MT, OR, LA, HI, NE, AR, MO, MA, KS, ID, AL, AZ, PR, CO, SC. OUS: IN, KR, SG, TH, VN, AR, BR, CA, CO, PA, UY, BE, CZ, DE, ES, FR, GB, IT, PT, RE, SE, SK, TM, JP.
Code information
UDI-DI: 08717648013591, Lots: 60418466, 60418467, 60418468