Z-0124-2021 Class II Terminated

Recalled by Quidel Corporation — San Diego, CA

FDA device recall Z-0124-2021 was initiated by Quidel Corporation on Sep 17, 2020 and is designated Class II. Reason for recall: Instructions for use in the package insert for SARS Antigen FIA test were updated to remove all references to a specific viral transport medium product as well as the use of other viral transport… The recall status is terminated (terminated Aug 9, 2022). Affected quantity: 10537.

Recall details

Product Type
Devices
Report Date
Oct 28, 2020
Initiation Date
Sep 17, 2020
Termination Date
Aug 9, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
10537

Product description

Sofia SARS Antigen FIA Package Insert The Sofia SARS Antigen FIA is a lateral flow immunofluorescent sandwich assay that is used with the Sofia and Sofia 2 instrument intended for the qualitative detection of the nucleocapsid protein antigen from SARS-CoV-2 in nasopharyngeal (NP) and nasal (NS) swab specimens directly from individuals who are suspected of COVID-19 by their healthcare provider within the first five days of the onset of symptoms.

Reason for recall

Instructions for use in the package insert for SARS Antigen FIA test were updated to remove all references to a specific viral transport medium product as well as the use of other viral transport media. Use of the specific viral transport medium products may result in False-positive results. Directly testing from specimens (NS or NP swab) is recommended.

Distribution pattern

U.S. Nationwide Distribution: NY, KS, TX, AL, FL, MN, CA, IL, MA, MS, PA, GA, NH, MO, VA, MI, OR, WV, OK, NV, CT, TN, MD, LA, AR, HI, NJ, KY, VT, IA, RI, NM, OH, IN, AK, SC, NC, AZ, SD, NE, WI, and.CO.

Code information

1438900EN00 (03/20) and 1438901EN00 (05/20)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0124-2021
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Quidel Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0124-2021. Report a data issue: hello@regdatalab.com.