Z-1086-2024 Class II Ongoing
FDA device recall Z-1086-2024 was initiated by Quidel Corporation on Jan 3, 2024 and is designated Class II. Reason for recall: One lot of the Sofia 2 Flu + SARS Antigen FIA (Lot 709083) was labelled with incorrect expiration dates. The kits contain pouched cassettes labeled with an expiration date of May 31, 2025, however the… The recall status is ongoing. Affected quantity: 312 Kits.
Recall details
- Product Type
- Devices
- Report Date
- Feb 21, 2024
- Initiation Date
- Jan 3, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 312 Kits
Product description
The Sofia 2 Flu + SARS Antigen FIA employs immunofluorescence technology in a sandwich design that is used with Sofia 2 to detect nucleocapsid protein from influenza A, influenza B, and SARS-CoV-2. This test allows for the detection of SARS-CoV and SARS-CoV-2, but does not differentiate between the two viruses.
Reason for recall
One lot of the Sofia 2 Flu + SARS Antigen FIA (Lot 709083) was labelled with incorrect expiration dates. The kits contain pouched cassettes labeled with an expiration date of May 31, 2025, however the correct expiration date for the pouches is September 12, 2024. The kits are labeled with an expiration date of April 30, 2025, however, the correct expiration date for the kits is September 12, 2024. The issue was discovered on November 28, 2023 by a supply chain employee who determined that the kits had expiration dates beyond the FDA authorized shelf life.
Distribution pattern
US Nationwide distribution in the states of FL, TN, IL, LA, NY, IN.
Code information
UDI-DI: 30014613339229. Lot: 709083
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1086-2024
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab