Z-2477-2021 Class II Terminated

Recalled by Quidel Corporation — Athens, OH

FDA device recall Z-2477-2021 was initiated by Quidel Corporation on Aug 13, 2021 and is designated Class II. Reason for recall: Revised Instructions for Use: Potential for specimens from patients with high SARS-CoV-2 viral loads to be falsely interpreted as negative if target amplification occurs prior to a cycle-threshold… The recall status is terminated (terminated Nov 2, 2023). Affected quantity: 81,910 kits.

Recall details

Product Type
Devices
Report Date
Sep 22, 2021
Initiation Date
Aug 13, 2021
Termination Date
Nov 2, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
81,910 kits

Product description

Lyra Direct SARS-CoV-2 Assay Emergency Use Authorization- Ref: M124. The Lyra Direct SARS-CoV-2 Assay is a real-time RT-PCR assay intended for the qualitative detection of nucleic acid from SARS-CoV-2 in nasal (NS), nasopharyngeal (NP), or oropharyngeal (OP) direct swab specimens from individuals suspected of COVID-19 by their healthcare provider

Reason for recall

Revised Instructions for Use: Potential for specimens from patients with high SARS-CoV-2 viral loads to be falsely interpreted as negative if target amplification occurs prior to a cycle-threshold (Ct) value of 5 when using the Thermo Fisher QuantStudio 7 Pro, the Applied Biosystems 7500 Fast Dx, the Applied Biosystems 7500, the Bio-Rad CFX96 Touch", the Roche LightCycler LC 480 II/ cobas z 480, or the Qiagen Rotor-Gene¿ MDx.

Distribution pattern

Worldwide distribution - US Nationwide distribution including Puerto Rico and the countries of Canada, Panama, Virgin Islands, Germany, Honduras.

Code information

All lots

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2477-2021
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2477-2021. Report a data issue: hello@regdatalab.com.