Z-0127-2024 Class II Ongoing

Recalled by Acclarent, Inc. — Irvine, CA

Recall Details

Product Type
Devices
Report Date
November 1, 2023
Initiation Date
September 21, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
141

Product Description

TruDi Navigation System, Model: FG-2000-00, Catalog: ENS022B, when used with TruDi Curette, Models: TDC0005Z and TDC0005

Reason for Recall

When using affected curette and software, there is a discrepancy between the actual curette tip location and the location displayed on navigation systems intended for use during surgical procedures of the Ear Nose and Throat (ENT) and ENT skull base surgery, which may cause delayed/prolonged surgery, cerebrospinal fluid leak, visual impairment, or skull base structural damage.

Distribution Pattern

US: CA, IL, WI, MA, TX, CO, FL, ID, KS, OH, OR, NE, MO, AZ, HI, TN, NY, ME, SC, NC, CT, DC, MN, LA, OK, PA, AL, IA, MI, SD

Code Information

TruDi Navigation System, UDI-DI: 10846835018639, software version: V2.3.1 Update (2.3.1.144 and 2.3.1.166)