Z-0825-2020 Class II Terminated
FDA device recall Z-0825-2020 was initiated by Acclarent, Inc. on Mar 6, 2019 and is designated Class II. Reason for recall: The firm became aware that three (3) lots of the NAV Suction Instruments contain calibration issues causing incorrect tip orientation display in the system. The recall status is terminated.
Recall details
- Product Type
- Devices
- Report Date
- Jan 29, 2020
- Initiation Date
- Mar 6, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
Product description
TruDi NAV Suction, TDNS070Z, 70 Degree Tip Angle (01) 10705031245884
Reason for recall
The firm became aware that three (3) lots of the NAV Suction Instruments contain calibration issues causing incorrect tip orientation display in the system.
Distribution pattern
US: WI, IL, NE, CA, OH, NY OUS: None
Code information
Lot # 1901080, 1901177, 1901219 The TruDi NAV Suction Instruments have a shelf life of 30 reprocessing cycles.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0825-2020
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026