Z-0415-2022 Class II Ongoing

Recalled by Acclarent, Inc. — Irvine, CA

Recall Details

Product Type
Devices
Report Date
December 29, 2021
Initiation Date
November 12, 2021
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1543

Product Description

TruDi NAV Suction Instruments: 0-Degree, REF: TDNS000Z; 70-Degree, REF: TDNS070Z; and 90-Degree, REF: TDNS090Z

Reason for Recall

Suction Instruments were incorrectly calibrated so they may not meet the system accuracy specification for visual verification of device location in the patient anatomy. If the surgeon does not detect the accuracy error, this could lead to an intraprocedural delay, a potential for reversible tissue damage or damage to the carotid artery, CSF leak, or damage to optic nerve.

Distribution Pattern

US: AZ, MI, IA, LA, IL, VA, SC, NY, IN, OR, SD, TX, PA, NE, FL, CT, MN, NV, CA, NJ, UT, AL, MA, WI, TN, WV, OH, NC, KY, HI¿, OK, GA, CO, MO, WY, MS, NH, HI, MD, ID

Code Information

0-Degree Lots: 1903120, 1903250, 1904060, 1904180, 1904240, 1905030, 1905070, 1906260, 1907120, 1907170, 1907220, 1907240, 1907260, 2003040, 2009110, 2011240, 2102080, 2103090, 2104260, 2105040, 2105260, 2106100, 2107070. 70-Degree Lots: 1903137, 1903257, 1904067, 1905017, 1905037, 1905077, 1906067, 1906267, 1907127, 1908137, 1908157, 2003127. 90-Degree: 1903139, 1903259, 1904069, 1905019, 1905139, 1906079, 1908269, 1909279, 1909309, 1910299, 2003029, 2004029