Z-0134-2025 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- October 30, 2024
- Initiation Date
- August 13, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1373 units
Product Description
Soltive Premium SuperPulsed Laser - Intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, Model/Catalog Number: TFL-PLS
Reason for Recall
The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of energy to the patients anatomy.
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of AU, CA, CL, DE, HK, JP, SG.
Code Information
UDI 00821925044111 All Serial Numbers: