Z-0143-2014 Class II Terminated

Recalled by Baxter Healthcare Corp. — Deerfield, IL

FDA device recall Z-0143-2014 was initiated by Baxter Healthcare Corp. on Sep 10, 2013 and is designated Class II. Reason for recall: The firm is recalling lots 10043 and 10044 due to loose particulate matter found in the packaging. Particulate matter entering the fluid path from the Luer lock caps may result in embolic events. The recall status is terminated (terminated Jun 5, 2015). Affected quantity: 628,992 units.

Recall Details

Product Type
Devices
Report Date
Nov 13, 2013
Initiation Date
Sep 10, 2013
Termination Date
Jun 5, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
628,992 units

Product Description

Dual Luer Lock Cap The device is intended for use as a cap for male or female Luer ports on medical devices such as manifolds, stopcocks or sets.

Reason for Recall

The firm is recalling lots 10043 and 10044 due to loose particulate matter found in the packaging. Particulate matter entering the fluid path from the Luer lock caps may result in embolic events.

Distribution Pattern

Worldwide Distribution - USA (nationwide) and the country of Canada.

Code Information

10043, 10044