Z-0170-2026 Class II Ongoing

Recalled by American Contract Systems Inc. — Grimes, IA

Recall Details

Product Type
Devices
Report Date
October 22, 2025
Initiation Date
September 4, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
32,433 total

Product Description

Basic Biopsy Tray, Item Number NMBP44L

Reason for Recall

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Distribution Pattern

US Nationwide distribution in the states of SD, IA, MN, WA, IL.

Code Information

UDI-DI 191072188120 Lot 8515411 Entire lot impacted