Z-0181-2026 Class II Ongoing

Recalled by American Contract Systems Inc. — Grimes, IA

Recall Details

Product Type
Devices
Report Date
October 22, 2025
Initiation Date
September 4, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
32,433 total

Product Description

Medical convenience kits Item Number/Description FHCP08Y C SECTION PK HICS50I C-SECTION PACK HSCB90B C-SECTION BASIN PACK

Reason for Recall

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Distribution Pattern

US Nationwide distribution in the states of SD, IA, MN, WA, IL.

Code Information

FHCP08Y UDI-DI 191072232724 Lot 8216211 Bag serial number 68258296; HICS50I UDI-DI 191072212351 Lot 8161711 Bag serial number 68273637; HSCB90B UDI-DI 191072141606 Lot 8083511 Bag serial numbers 47506618 47508489 * If bag serial number is listed as "N/A", then whole lot is affected