Z-0176-2026 Class II Ongoing

Recalled by American Contract Systems Inc. — Grimes, IA

Recall Details

Product Type
Devices
Report Date
October 22, 2025
Initiation Date
September 4, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
32,433 total

Product Description

Medical convenience kits Item Number/Description ANCV78BC HEART PACK - 205947 HGCV05K CV SET UP PACK MHPP97AR PACEMAKER PACK- 206010 UDBH37AW BASIC HEART PACK - 206044 UHPP82AH PERIPHERAL PACK - 209533 UICD62AN CARDIOVASCUALR SUPPLY (PS 907285) UICT16F CARDIO THORACIC DRAPE PACK (PS 907284) UIVA75I VASCULAR PACK

Reason for Recall

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Distribution Pattern

US Nationwide distribution in the states of SD, IA, MN, WA, IL.

Code Information

ANCV78BB UDI-DI 191072214362 Lot 8133811 Bag serial number 68273380; ANCV78BC UDI-DI 191072235169 Lot 8499811 Bag serial numbers N/A; HGCV05K UDI-DI 191072212474 Lot 8021411 Bag serial number 68154464; MHPP97AR UDI-DI 191072237446 Lot 8527511 Bag serial number N/A; UDBH37AW UDI-DI 191072235299 Lot 8194311 Bag serial numbers 68198334 68198335; UHPP82AH UDI-DI 191072216434 Lot 8112811 Bag serial number N/A; UICD62AN UDI-DI 191072200495 Lot 7999011 Bag serial number N/A; UICT16F UDI-DI 191072150431 Lot 8000411 Bag serial number N/A Lot 8331311 Bag serial number 17627819 Lot 8429011 Bag serial number 17749736; UIVA75I UDI-DI 191072229052 Lot 8056011 Bag serial number N/A; UIVA75J UDI-DI 191072235602 Lot 8422011 Bag serial number 68667536 * If Bag serial number is "n/a", then all of that lot is affected