Z-0178-2026 Class II Ongoing

Recalled by American Contract Systems Inc. — Grimes, IA

FDA device recall Z-0178-2026 was initiated by American Contract Systems Inc. on Sep 4, 2025 and is designated Class II. Reason for recall: Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for… The recall status is ongoing. Affected quantity: 32,433 total.

Recall details

Product Type
Devices
Report Date
Oct 22, 2025
Initiation Date
Sep 4, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
32,433 total

Product description

Medical convenience kits Item Number/Description RCEN25B ENT SDSC 0246972 RCHN27D HEAD AND NECK RCH 0246951

Reason for recall

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Distribution pattern

US Nationwide distribution in the states of SD, IA, MN, WA, IL.

Code information

RCEN25B UDI-DI 191072225115 lot 8138311 bag serial number N/A; RCHN27D UDI-DI 191072235923 lot 8508211 bag serial number 47847459 * If bag serial number is listed as "n/a", then entire lot is affected

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0178-2026
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0178-2026. Report a data issue: hello@regdatalab.com.