Z-0182-2026 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- October 22, 2025
- Initiation Date
- September 4, 2025
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 32,433 total
Product Description
Medical convenience kits Item Number/Description LKRA40G RADIOLOGY PACK
Reason for Recall
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Distribution Pattern
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
Code Information
LKRA40G UDI-DI 191072189691 Lot 8025211 Bag serial number 68154347 * If bag serial number is listed as "N/A", then whole lot is affected