Z-0198-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- November 16, 2022
- Initiation Date
- September 29, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 65
Product Description
Canon Aplio ultrasound system, Model: Aplio i900 TUS-AI900, Aplio i800 TUS-AI800, Aplio i700 TUS-AI700, Aplio a550 CUS-AA550, Aplio a450 CUS-AA450
Reason for Recall
After completion, ultrasound system's Stress Echo may not terminate if user does not manually terminate the protocol. During next patient exam, an abnormal confirmation message will be displayed "Same view was saved in this phase. Do you use this image with StressEcho Review?" If "no", acquired images not saved, and if "yes" images assigned to the previous patient; could cause treatment delay.
Distribution Pattern
US: MT, MO, NC, SC, MN, WA, KY, TN, AR, IL, IA, MI, WV, KS, CA, HA, CO, LA, OK, ME, KY, NY, PA, GA, FL
Code Information
Model/UDI-DI, with software: Aplio i900 TUS-AI900/ 04987670102523, Aplio i800 TUS-AI800/ 04987670102516, Aplio i700 TUS-AI700/ 04987670102509, With System Version:V4.6/V5.0/V5.1, SP0005 to SP1014, or V6.0/V6.5 SP0000 to SP1009; Aplio a550 CUS-AA550/ 04987670103353, Aplio a450 CUS-AA450/ 04987670103346, With System Version: V3.0, SP0005 to SP1014, or V4.0/V4.5, SP0000 to SP1009