Z-0198-2023 Class II Ongoing

Recalled by Canon Medical System, USA, INC. — Tustin, CA

Recall Details

Product Type
Devices
Report Date
November 16, 2022
Initiation Date
September 29, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
65

Product Description

Canon Aplio ultrasound system, Model: Aplio i900 TUS-AI900, Aplio i800 TUS-AI800, Aplio i700 TUS-AI700, Aplio a550 CUS-AA550, Aplio a450 CUS-AA450

Reason for Recall

After completion, ultrasound system's Stress Echo may not terminate if user does not manually terminate the protocol. During next patient exam, an abnormal confirmation message will be displayed "Same view was saved in this phase. Do you use this image with StressEcho Review?" If "no", acquired images not saved, and if "yes" images assigned to the previous patient; could cause treatment delay.

Distribution Pattern

US: MT, MO, NC, SC, MN, WA, KY, TN, AR, IL, IA, MI, WV, KS, CA, HA, CO, LA, OK, ME, KY, NY, PA, GA, FL

Code Information

Model/UDI-DI, with software: Aplio i900 TUS-AI900/ 04987670102523, Aplio i800 TUS-AI800/ 04987670102516, Aplio i700 TUS-AI700/ 04987670102509, With System Version:V4.6/V5.0/V5.1, SP0005 to SP1014, or V6.0/V6.5 SP0000 to SP1009; Aplio a550 CUS-AA550/ 04987670103353, Aplio a450 CUS-AA450/ 04987670103346, With System Version: V3.0, SP0005 to SP1014, or V4.0/V4.5, SP0000 to SP1009