Z-0207-2019 Class II Terminated
FDA device recall Z-0207-2019 was initiated by Howmedica Osteonics Corp. on Sep 12, 2018 and is designated Class II. Reason for recall: The firm received complaints related to the locking mechanism on the steerable inserter not functioning as intended. The firm confirmed that the issue is related to lack of lubrication of the… The recall status is terminated (terminated Jul 16, 2020). Affected quantity: 58.
Recall Details
- Product Type
- Devices
- Report Date
- Oct 31, 2018
- Initiation Date
- Sep 12, 2018
- Termination Date
- Jul 16, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 58
Product Description
Tritanium TL Steerable Inserter, Cat. No. 48930000, UDI 07613327381900 Product Usage: The TL Steerable Inserter is a Class 1 instrument which is provided in a non-sterile condition. The inserter is designed to enter the disc space, with a fixed engagement between the inserter and the cage and then rotate for desired placement. The three components of the TL Steerable Inserter System include: 1. TL Steerable Inserter 2. TL Steerable Inserter Inner Shaft 3. TL Steerable Inserter Driver
Reason for Recall
The firm received complaints related to the locking mechanism on the steerable inserter not functioning as intended. The firm confirmed that the issue is related to lack of lubrication of the instrument. There are two potential hazards associated with the implant pivoting unexpectedly during implant insertion into the disc space. First, if the implant is initially inserted into the disc space via the impaction method and the implant pivots unexpectedly, the implant potentially may unintentionally contact sensitive anatomy. Additionally, the user may potentially experience difficulty positioning the implant within the disc space due to the lack of rigidity. A third potential hazard is associated with the inserter not unlocking at the articulating pivot head during implant positioning within the disc space. Once the implant is in the prepared disc space and contacts the ventral annulus, the surgeon unlocks the articulating pivot head and positions the implant perpendicular to the spinous process. The surgeon may then elect to lock/unlock the articulating pivot head for desired final positioning of the implant. If the inserter does not unlock during these final implant positioning steps, the user may potentially implant the cage in a suboptimal trajectory/placement.
Distribution Pattern
US Nationwide Distribution.
Code Information
All lots