Z-3204-2026 Class II Ongoing

Recalled by Howmedica Osteonics Corp. — Mahwah, NJ

FDA device recall Z-3204-2026 was initiated by Howmedica Osteonics Corp. on July 30, 2026 and is designated Class II. Reason for recall: Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which m… The recall status is ongoing. Affected quantity: 12,156 US.

Recall Details

Product Type
Devices
Report Date
September 23, 2026
Initiation Date
July 30, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
12,156 US

Product Description

TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Reason for Recall

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.

Code Information

620-00-28D UDI-DI 07613327313581 Lots MHANP2 MHD1N0