Z-0265-2025 Class II Ongoing

Recalled by Beckman Coulter, Inc. — Chaska, MN

FDA device recall Z-0265-2025 was initiated by Beckman Coulter, Inc. on Sep 5, 2024 and is designated Class II. Reason for recall: IFUs for the device incorrectly states the biotin interference was remediated. The issue could result in potentially false-low results. The recall status is ongoing. Affected quantity: 465,481 kits.

Recall Details

Product Type
Devices
Report Date
Nov 6, 2024
Initiation Date
Sep 5, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
465,481 kits

Product Description

Access Thyroglobulin Antibody II, REF A32898, containing Instructions for Use, IFU REF B14994, B63978, and A38168. The Access Thyroglobulin Antibody II (TgAb) assay is used for the quantitative determination of thyroglobulin antibody levels in human serum and plasma using the Access Immunoassay Systems.

Reason for Recall

IFUs for the device incorrectly states the biotin interference was remediated. The issue could result in potentially false-low results.

Distribution Pattern

Distribution was made nationwide to the U.S., including Puerto Rico. There was government distribution but no military distribution. Foreign distribution was made to Mexico.

Code Information

All IFU Part Numbers B14994, B63978, and A38168 affect all product/lots shipped from Jul-2021 to current; UDI-DI 15099590202910.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0265-2025
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.