Z-0268-2016 Class II Terminated

Recalled by Medtronic Perfusion Systems — Brooklyn Park, MN

FDA device recall Z-0268-2016 was initiated by Medtronic Perfusion Systems on Oct 28, 2015 and is designated Class II. Reason for recall: This recall is being initiated due to a potential breach of the Tubing Pack Header Bags manufactured in these select lots. The seal in question is between the Tyvek and the Poly side of the header… The recall status is terminated (terminated Aug 17, 2016). Affected quantity: 496.

Recall Details

Product Type
Devices
Report Date
Nov 18, 2015
Initiation Date
Oct 28, 2015
Termination Date
Aug 17, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
496

Product Description

Medtronic Perfusion Tubing Packs, Models TL7M64R3, TL7P17R1 & HY2W85R9, Sterilized using Ethylene Oxide. The Tubing Pack design and components are specified by the user, they are packaged with consideration of setup efficiency and respect for the environment, and they are intended for use in the extracorporeal circuit during cardiopulmonary bypass procedures. Medtronic Tubing Packs can contain components with Carmeda BioActive Surface, Trillium BioSurface, Balance BioSurface or no coating. Sterile barrier for these packs is provided either by a Tyvek lid seal on the tray or by a header bag.

Reason for Recall

This recall is being initiated due to a potential breach of the Tubing Pack Header Bags manufactured in these select lots. The seal in question is between the Tyvek and the Poly side of the header bag.

Distribution Pattern

Nationwide Distribution including US: AL, IL, IN, and WI.

Code Information

Model Lot Numbers Lot Number HY2W85R9 209704344,209704345,209704346, 209704347,210010558 TL7M64R3 209789393 TL7P17R1 209789333

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0268-2016
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.