Z-0284-2021 Class II Terminated

Recalled by Philips North America, LLC — Andover, MA

FDA device recall Z-0284-2021 was initiated by Philips North America, LLC on Sep 14, 2020 and is designated Class II. Reason for recall: The periodic Paddle Checks recommended in the Instructions for Use for Sterilizable Defibrillator Paddles may not detect one failure mode for the Switched Internal Defibrillator Paddles. Philips has… The recall status is terminated (terminated Sep 13, 2022). Affected quantity: 8,942.

Recall details

Product Type
Devices
Report Date
Nov 4, 2020
Initiation Date
Sep 14, 2020
Termination Date
Sep 13, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8,942

Product description

Philips Sterilizable Defibrillator Paddles, Switched Internal Paddles, Model Numbers M4741A, M4742A, M4743A, M4744A - Product Usage: used in pairs to transmit a controlled electrical shock from an external defibrillator directly to the exposed heart muscle of a post-pubescent patient in order to intentionally stop start the heartbeat during cardiopulmonary surgery.

Reason for recall

The periodic Paddle Checks recommended in the Instructions for Use for Sterilizable Defibrillator Paddles may not detect one failure mode for the Switched Internal Defibrillator Paddles. Philips has created an addendum to the IFU, that includes a test for the switched paddles only, which will detect this failure mode.

Distribution pattern

Global Distribution. US Nationwide.

Code information

All units manufactured and distributed January 2015 to August 2020.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0284-2021
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips North America, LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips North America, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips North America, LLC →

Source: openFDA record for Z-0284-2021. Report a data issue: hello@regdatalab.com.