Z-1963-2026 Class II Ongoing

Recalled by Philips North America — Cambridge, MA

FDA device recall Z-1963-2026 was initiated by Philips North America on Apr 14, 2026 and is designated Class II. Reason for recall: The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings… The recall status is ongoing. Affected quantity: 103 units.

Recall details

Product Type
Devices
Report Date
May 6, 2026
Initiation Date
Apr 14, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
103 units

Product description

Philips Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 781358. 2. Model Number (REF): 782107. 3. Model Number (REF): 782136. 510(k) Numbers: K213516, K223458, K230972, K232030, K251397, K251808

Reason for recall

The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

Distribution pattern

Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar, Netherlands, Pakistan, Palestine, Peru, Poland, Portugal, Russian Fed., Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Kingdom.

Code information

1. Model Number (REF): 781358. UDI: (01)00884838088115(21)45121, (01)00884838088115(21)45148, (01)00884838088115(21)45160, (01)00884838088115(21)45438, (01)00884838088115(21)45034, (01)00884838088115(21)45049, (01)00884838088115(21)45066, (01)00884838088115(21)45090, (01)00884838088115(21)45141, (01)00884838088115(21)45149, (01)00884838088115(21)45170, (01)00884838088115(21)45250, (01)00884838088115(21)45290, (01)00884838088115(21)45085, (01)00884838088115(21)45312, (01)00884838088115(21)45376, (01)00884838088115(21)45122, (01)00884838088115(21)45169, (01)00884838088115(21)45191, (01)00884838088115(21)45192, (01)00884838088115(21)45203, (01)00884838088115(21)45204, (01)00884838088115(21)45207, (01)00884838088115(21)45220, (01)00884838088115(21)45228, (01)00884838088115(21)45229, (01)00884838088115(21)45249, (01)00884838088115(21)45278, (01)00884838088115(21)45279, (01)00884838088115(21)45320, (01)00884838088115(21)45028, (01)00884838088115(21)45195, (01)00884838088115(21)45174, (01)00884838088115(21)45165, (01)00884838088115(21)45382, (01)00884838088115(21)45011, (01)00884838088115(21)45055, (01)00884838088115(21)45075, (01)00884838088115(21)45113, (01)00884838088115(21)45147, (01)00884838088115(21)45152, (01)00884838088115(21)45235, (01)00884838088115(21)45237, (01)00884838088115(21)45245, (01)00884838088115(21)45264, (01)00884838088115(21)45269, (01)00884838088115(21)45277, (01)00884838088115(21)45316, (01)00884838088115(21)45322, (01)00884838088115(21)45337, (01)00884838088115(21)45338, (01)00884838088115(21)45353, (01)00884838088115(21)45398, (01)00884838088115(21)45414, (01)00884838088115(21)45429, (01)00884838088115(21)45051. Serial Numbers: 45121, 45148, 45160, 45438, 45034, 45049, 45066, 45090, 45141, 45149, 45170, 45250, 45290, 45085, 45312, 45376, 45122, 45169, 45191, 45192, 45203, 45204, 45207, 45220, 45228, 45229, 45249, 45278, 45279, 45320, 45028, 45195, 45174, 45165, 45382, 45011, 45055, 45075, 45113, 45147, 45152, 45235, 45237, 45245, 45264, 45269, 45277, 45316, 45322, 45337, 45338, 45353, 45398, 45414, 45429, 45051. 2. Model Number (REF): 782107. UDI: (01)00884838098336(21)45590, (01)00884838098336(21)62158, (01)00884838098336(21)62180, (01)00884838098336(21)62044, (01)00884838098336(21)45567, (01)00884838098336(21)62031, (01)00884838098336(21)62166, (01)00884838098336(21)62184, (01)00884838098336(21)62190, (01)00884838098336(21)62240, (01)00884838098336(21)45519, (01)00884838098336(21)45533, (01)00884838098336(21)62024, (01)00884838098336(21)62047, (01)00884838098336(21)62048, (01)00884838098336(21)62051, (01)00884838098336(21)62052, (01)00884838098336(21)45554, (01)00884838098336(21)62145, (01)00884838098336(21)62146, (01)00884838098336(21)62148, (01)00884838098336(21)45500, (01)00884838098336(21)45501, (01)00884838098336(21)45510, (01)00884838098336(21)45511, (01)00884838098336(21)45514, (01)00884838098336(21)45516, (01)00884838098336(21)45518, (01)00884838098336(21)45589, (01)00884838098336(21)62017, (01)00884838098336(21)62039, (01)00884838098336(21)62073, (01)00884838098336(21)62081, (01)00884838098336(21)62088, (01)00884838098336(21)45521, (01)00884838098336(21)45600, (01)00884838098336(21)45563, (01)00884838098336(21)45570, (01)00884838098336(21)62153, (01)00884838098336(21)62181, (01)00884838098336(21)62182, (01)00884838098336(21)62183. Serial Numbers: 45590, 62158, 62180, 62044, 45567, 62031, 62166, 62184, 62190, 62240, 45519, 45533, 62024, 62047, 62048, 62051, 62052, 45554, 62145, 62146, 62148, 45500, 45501, 45510, 45511, 45514, 45516, 45518, 45589, 62017, 62039, 62073, 62081, 62088, 45521, 45600, 45563, 45570, 62153, 62181, 62182, 62183. 3. Model Number (REF): 782136. UDI: (01)00884838108608(21)28568, (01)00884838108608(21)28503, (01)00884838108608(21)28515, (01)00884838108608(21)28528, (01)00884838108608(21)28577. Serial Numbers: 28568, 28503, 28515, 28528, 28577.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1963-2026
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips North America? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1963-2026. Report a data issue: hello@regdatalab.com.