Z-1960-2026 Class II Ongoing

Recalled by Philips North America — Cambridge, MA

Recall Details

Product Type
Devices
Report Date
May 6, 2026
Initiation Date
April 14, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7 units

Product Description

Philips Ingenia Ambition S with MR Elastography (MRE). 1. Model Number (REF): 781359. 2. Model Number (REF): 782108. 3. Model Number (REF): 782139.

Reason for Recall

The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

Distribution Pattern

Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar, Netherlands, Pakistan, Palestine, Peru, Poland, Portugal, Russian Fed., Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Kingdom.

Code Information

1. Model Number (REF): 781359. UDI: (01)00884838090057(21)48218, (01)00884838090057(21)48233. Serial Numbers: 48218, 48233. 2. Model Number (REF): 782108. UDI: (01)00884838098343(21)48652, (01)00884838098343(21)48551, (01)00884838098343(21)48669. Serial Numbers: 48652, 48551, 48669. 3. Model Number (REF): 782139. UDI: (01)00884838108639(21)29076, (01)00884838108639(21)29079. Serial Numbers: 29076, 29079.