Z-0285-2019 Class II Terminated

Recalled by Magellan Diagnostics, Inc. — North Billerica, MA

Recall Details

Product Type
Devices
Report Date
November 7, 2018
Initiation Date
August 8, 2018
Termination Date
May 12, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,465 kits

Product Description

Magellan LeadCare¿ II Blood Lead Test Kit. In Vitro Test for the measurement of lead in fresh whole blood. Catalog Number: 70-6762 Primary DI No: 00850355006000 The LeadCare II Blood Lead Test Kit is for in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in fresh whole blood. This product is for professional use only. Sensors and treatment reagent of the test kit are specific for lead only. Contents of the LeadCare II Blood Lead Test Kit must only be used with a LeadCare II Analyzer.

Reason for Recall

Defect with the calibration button for lot 1808M. Level 1 and Level 2 Quality Control (QC) values will generate results of HIGH (>65 ¿g/dL). Results for patient testing with this lot will be displayed as HIGH.

Distribution Pattern

Worldwide distribution: US (Nationwide) and countries of: Spain and South Africa.

Code Information

Lot 1808M and sub-lots: 1808M-01, 1808M-02, 1808M-03, 1808M-04, 1808M-05, 1808M-06, 1808M-07, 1808M-08, 1808M-09. Exp. Date:November 10, 2019