Z-1953-2021 Class I Terminated
FDA device recall Z-1953-2021 was initiated by Magellan Diagnostics, Inc. on May 7, 2021 and is designated Class I. Reason for recall: Magellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient… The recall status is terminated (terminated Oct 24, 2023). Affected quantity: 320 kits (96 tests/kit=30,720 tests).
Recall details
- Product Type
- Devices
- Report Date
- Jul 7, 2021
- Initiation Date
- May 7, 2021
- Termination Date
- Oct 24, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 320 kits (96 tests/kit=30,720 tests)
Product description
Reason for recall
Distribution pattern
Worldwide distribution. US nationwide including Puerto Rico; Canada, Vietnam, India, Italy, Peru, New Zealand, Australia, Hong Kong, and Philippines.