Z-0219-2024 Class II Ongoing

Recalled by Magellan Diagnostics, Inc. — North Billerica, MA

FDA device recall Z-0219-2024 was initiated by Magellan Diagnostics, Inc. on Sep 19, 2023 and is designated Class II. Reason for recall: During an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may have inadequate… The recall status is ongoing. Affected quantity: 11,442 LeadCare II kits (549,216 tests).

Recall details

Product Type
Devices
Report Date
Nov 15, 2023
Initiation Date
Sep 19, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
11,442 LeadCare II kits (549,216 tests)

Product description

LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in capillary whole blood.

Reason for recall

During an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may have inadequate cap-to-tube seals. Manufacturing with these caps may result in cap-to-tube seals that allow leaking of HCl or may result in cap-to-tube seals that allow evaporation of treatment reagent. The consequence of this would be a change in the volume of treatment reagent and a change in the ratios of either added patient sample or control material to treatment reagent. The result of this might be a falsely elevated result (e.g., a patient false positive [PATFP] result or a Controls Out of Range-High [COOR-Hi] result).

Distribution pattern

US Nationwide distribution.

Code information

UDI-DI :(01)00850355006000 Lot Numbers: 2231 2231-10 2303M-11 2303M-12 2303M-13 2303M-15 2305M-01 2305M-02 2305M-03 2305M-04 2305M-05 2306M-01 2306M-02 2306M-03 2306M-04 2306M-05 2306M-06 2307M 2307M-04 2307M-05 2307M-06 2307M-07 2307M-10 2307M-11 2308M-01 2308M-02 2308M-03 2308M-07 2309M 2310M-01 2310M-03 2310M-04 2310M-05 2310M-06 2310M-08 2311M-02 2311M-03 2311M-05 2311M-09 2311M-10 2312M 2312M-01 2312M-04 2312M-05 2312M-07 2313M 2313M-01 2313M-05 2313M-07 2314M-06 2314M-07 2314M-08 2314M-09 2314M-12 2314M-13 2315M-02 2315M-03 2315M-04 2315M-05 2315M-06 2315M-07 2315M-08 2315M-13 2316M 2316M-03 2316M-04 2316M-05 2316M-06 2316M-11 2316M-12 2317M 2317M-01 2318M 2320M-06 7308M

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0219-2024
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Magellan Diagnostics, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0219-2024. Report a data issue: hello@regdatalab.com.