Z-0285-2025 Class II Ongoing

Recalled by Zimmer, Inc. — Warsaw, IN

FDA device recall Z-0285-2025 was initiated by Zimmer, Inc. on Sep 17, 2024 and is designated Class II. Reason for recall: A Gen 5 tail piece was assembled to a Gen 4 Finished Device, resulting in a 0.178" difference in the end of the endoscope slot to the end of the tail piece. This allows the scope to move further into… The recall status is ongoing. Affected quantity: 432 units.

Recall Details

Product Type
Devices
Report Date
Nov 13, 2024
Initiation Date
Sep 17, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
432 units

Product Description

Saphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally invasive surgery allowing access for vessel harvesting and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. Item Number: VPX4000 (7273)

Reason for Recall

A Gen 5 tail piece was assembled to a Gen 4 Finished Device, resulting in a 0.178" difference in the end of the endoscope slot to the end of the tail piece. This allows the scope to move further into the plastic sheath that runs the length of the device, resulting in an interference fit at the cone, may caus an extension in surgery

Distribution Pattern

Nationwide

Code Information

UDI: +M718VPX40001/$$326051543DD6738Q, +M718VPX4001/$$326062143DD8092J Lot Numbers: 43DD6738, 43DD8092

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0285-2025
openFDA file date
Oct 6, 2026
Loaded by RegDataLab
Oct 12, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it. FDA may have changed the recall since.