Z-0974-2026 Class II Ongoing
FDA device recall Z-0974-2026 was initiated by Zimmer, Inc. on Dec 2, 2025 and is designated Class II. Reason for recall: Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse… The recall status is ongoing. Affected quantity: 11.
Recall details
- Product Type
- Devices
- Report Date
- Dec 31, 2025
- Initiation Date
- Dec 2, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 11
Product description
Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609300 Software Version: N/A Product Description: Affixus 9 mm, Length 300 mm, Left, Trochanteric Nail Component: N/A
Reason for recall
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
Distribution pattern
US Nationwide distribution in the states of AL, AZ, CA, FL, IN, KY, MA, MO, MT, NC, NM, RI, TX, UT.
Code information
Lot Code: Model No 815609300 ; UDI-DI (01)00887868144462(17)340826(10)66717600 ; Lot Number 66717600