Z-0975-2026 Class II Ongoing

Recalled by Zimmer, Inc. — Warsaw, IN

FDA device recall Z-0975-2026 was initiated by Zimmer, Inc. on Dec 2, 2025 and is designated Class II. Reason for recall: Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse… The recall status is ongoing. Affected quantity: 23.

Recall details

Product Type
Devices
Report Date
Dec 31, 2025
Initiation Date
Dec 2, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
23

Product description

Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609320 Software Version: N/A Product Description: Affixus 9 mm, Length 320 mm, Left, Trochanteric Nail Component: N/A

Reason for recall

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.

Distribution pattern

US Nationwide distribution in the states of AL, AZ, CA, FL, IN, KY, MA, MO, MT, NC, NM, RI, TX, UT.

Code information

Lot Code: Model No 815609320 ; UDI-DI (01)00887868144479(17)340826(10)66717601 ; Lot Number 66717601 Model No 815609320 ; UDI-DI (01)00887868144479(17)340820(10)66892900 ; Lot Number 66892900

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0975-2026
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Zimmer, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0975-2026. Report a data issue: hello@regdatalab.com.