Z-0289-2013 Class III Terminated

Recalled by Horiba Instruments, Inc Dba Horiba Medical — Irvine, CA

FDA device recall Z-0289-2013 was initiated by Horiba Instruments, Inc Dba Horiba Medical on Sep 14, 2012 and is designated Class III. Reason for recall: This recall was initiated because some customers were experiencing an alarm during Creatinine calibration "REAG RANGE HIGH". This alarm caused the calibration to fail and prevented the user from using… The recall status is terminated (terminated Nov 29, 2012). Affected quantity: 8,503 units.

Recall details

Product Type
Devices
Report Date
Nov 21, 2012
Initiation Date
Sep 14, 2012
Termination Date
Nov 29, 2012
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8,503 units

Product description

ABX PENTRA Creatinine 120 CP, Model Number Creatinine Part # A11A01868, Lot Numbers: 203273706, 203273707, 203273808, 203273709. The ABX PENTRA Creatinine 120 CP is an in vitro diagnostic assay for the quantitative determination of creatinine in human serum, plasma and urine based on a kinetic method using alkaline picrate (Jaffe method). It is composed of a 27 ml monoreagent cassette. Reagent is a chemical solution with additives. ABX PENTRA Creatinine 120 CP reagent, with associated calibrator and controls, is a diagnostic reagent for quantitative in vitro determination of Creatinine in human serum, plasma and urine based on a kinetic method using alkaline picrate (Jaffe method). Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. This reagent is intended for use on the ABX PENTRA 400 Clinical Chemistry Analyzer.

Reason for recall

This recall was initiated because some customers were experiencing an alarm during Creatinine calibration "REAG RANGE HIGH". This alarm caused the calibration to fail and prevented the user from using the Creatinine assay.

Distribution pattern

Nationwide Distribution

Code information

Model Number Creatinine Part # A11A01868, Lot Numbers: 203273706, 203273707, 203273808, 203273709.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)