Z-0291-2019 Class II Terminated

Recalled by Stryker GmbH — Selzach, N/A

FDA device recall Z-0291-2019 was initiated by Stryker GmbH on Aug 28, 2018 and is designated Class II. Reason for recall: The press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced… The recall status is terminated (terminated Feb 4, 2021). Affected quantity: 867.

Recall Details

Product Type
Devices
Report Date
Nov 7, 2018
Initiation Date
Aug 28, 2018
Termination Date
Feb 4, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
867

Product Description

Hoffman LRF Hexapod Strut, Extra Short 183mm-295mm, Model Number 4935-0-040. Product Usage - The Hoffmann LRF System is indicated in pediatric patients and adults for the treatment and fixation of: Open and closed fractures Post-traumatic joint contracture which has resulted in loss of range of motion Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction Pseudoarthrosis or non-union of long bones Limb lengthening by epiphyseal, diaphyseal, or metaphyseal distraction Correction of bony or soft tissue deformity Correction of segmental bony or soft tissue defects Joint arthrodesis Management of communicated intra-articular fractures of the distal radius Bone transport The Hoffmann LRF System is indicated in adults for: Osteotomy Revision procedure where other treatments or devices have been unsuccessful Bone reconstruction procedures Fusions and replantations of the foot Charcot foot reconstruction Lisfranc dislocations The typical LRF construct features two or more rings bridged by three to four telescopic struts with multiplanar, ball jointed fixation bolts. Once applied, the LRF can be unlocked for gross frame adjustment and fracture reduction. The LRF can be finely adjusted to aid in limb alignment and definitively locked to hold compression and reduction. If deemed appropriate, the LRF Telescopic Struts can provide up to 5mm of controlled dynamization. Original Statement OP-Tech H-ST-1_Rev2_HoffmannLRF_OpTech_2016-10690]

Reason for Recall

The press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced press fit of the pin. This can result in partial of full displacement or loss of the pin and, accordingly, stability of the strut of the frame cannot be ensured.

Distribution Pattern

US Nationwide in the states of CO, FL, MD, MI, NY, OK, and PA.

Code Information

Lots: L18239 L13690 L22250 L26334 L31272 L35847 L35848 L46136 L46137 L46495 L46497 L50179 L51363 L53107 L53490 L59090 Y02173 Y07632 Y08556 Y08558 Y08557 Y14369 Y14370 Y16719 Y14646 Y14642

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0291-2019
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 24, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.